Performance of a monolithic LaBr3:Ce crystal coupled to an array of silicon photomultipliers

NázevTitle
Performance of a monolithic LaBr3:Ce crystal coupled to an array of silicon photomultipliersPerformance of a monolithic LaBr3:Ce crystal coupled to an array of silicon photomultipliers
Druh výsledkuResult type
Článek v časopiseJournal article
AutořiAuthors
A. Ulyanov, O. Morris, T. Slavíček, C. Granja, M. Solar
Klíčová slovaKeywords
Gamma-ray detector, scintillator, LaBr3, Lanthanum bromide, SiPM, silicon photomultiplier
DOIDOI
10.1016/j.nima.2015.11.148
Časopis / citaceJournal / citation
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 810, 107-119 (2016) · ISSN 0168-9002
RokYear
2016
JazykLanguage
eng
ZáznamyRecords
ProjektProject
Van de Graaff - urychlovač iontů HV2500 jako laditelný zdroj neutronů v rámci české a evropské velké infrastrukturyAccelerator of ions HV2500 as tuneable neutron source in frame of Czech and European Large Infrastructure
CitovánoCited by
31 (OpenAlex)
2016: 42017: 52018: 12019: 22020: 42021: 62022: 42023: 12024: 4
Citace ke staženíDownload citation
TXT · BibTeX

AbstraktAbstract

A gamma-ray detector composed of a single 28 x 28 x 20 mm(3) LaBr3:Ce crystal coupled to a custom built 4 x 4 array of silicon photomultipliers was tested over an energy range of 30 keV to 9.3 MeV. The silicon photomultipliers were initially calibrated using 20 ns light pulses generated by a light emitting diode. The photodetector responses measured as a function of the number of incident photons were found to be non-linear and consistent with model predictions. Using corrections for the non-linearity of the silicon photomultipliers, the detector showed a linear response to gamma-rays with energies from 100 keV to the maximum available energy of 9.3 MeV. The energy resolution was found to be 4% FWHM at 662 keV. Despite the large thickness of the scintillator (20 mm) and a 5 mm thick optical window, the detector was capable of measuring the positions of the gamma-ray interaction points. The position resolution was measured at 356 keV and was found to be 8 mm FVVHM in the detector plane and 11 mm FWHM for the depth of interaction. The detector can be used as a building block of a larger calorimeter system that is capable of measuring gamma-ray energies up to tens of MeV.

A gamma-ray detector composed of a single 28 x 28 x 20 mm(3) LaBr3:Ce crystal coupled to a custom built 4 x 4 array of silicon photomultipliers was tested over an energy range of 30 keV to 9.3 MeV. The silicon photomultipliers were initially calibrated using 20 ns light pulses generated by a light emitting diode. The photodetector responses measured as a function of the number of incident photons were found to be non-linear and consistent with model predictions. Using corrections for the non-linearity of the silicon photomultipliers, the detector showed a linear response to gamma-rays with energies from 100 keV to the maximum available energy of 9.3 MeV. The energy resolution was found to be 4% FWHM at 662 keV. Despite the large thickness of the scintillator (20 mm) and a 5 mm thick optical window, the detector was capable of measuring the positions of the gamma-ray interaction points. The position resolution was measured at 356 keV and was found to be 8 mm FVVHM in the detector plane and 11 mm FWHM for the depth of interaction. The detector can be used as a building block of a larger calorimeter system that is capable of measuring gamma-ray energies up to tens of MeV.