Ústav technické a experimentální fyziky Institute of Experimental and Applied Physics

First tests of Timepix detectors based on semi-insulating GaAs matrix of different pixel size

NázevTitle
First tests of Timepix detectors based on semi-insulating GaAs matrix of different pixel sizeFirst tests of Timepix detectors based on semi-insulating GaAs matrix of different pixel size
Druh výsledkuResult type
Článek v časopiseJournal article
AutořiAuthors
B. Zat'ko, D. Kubanda, J. Žemlička, A. Sagatova, Y. Mora, M. Pichotka, J. Dudák
DOIDOI
10.1088/1748-0221/13/02/C02013
Časopis / citaceJournal / citation
Journal of Instrumentation. 2018, 13 ISSN 1748-0221.
RokYear
2018
JazykLanguage
eng
WoSWoS
000424627200003
ScopusScopus
2-s2.0-85043477632
RIVRIV
RIV/68407700:21670/18:00319853!RIV19-MSM-21670___
ProjektProject
Urychlovač Van de Graaff - laditelný zdroj monoenergetických neutronů a lehkých iontůVan de Graaff Accelerator - a Tunable Source of Monoenergetic Neutrons and Light Ions

AbstraktAbstract

In this work, we have focused on Timepix detectors coupled with the semi-insulating GaAs material sensor. We used undoped bulk GaAs material with the thickness of 350 mu m. We prepared and tested four pixelated detectors with 165 mu m and 220 mu m pixel size with two versions of technology preparation, without and with wet chemically etched trenches around each pixel. We have carried out adjustment of GaAs Timepix detectors to optimize their performance. The energy calibration of one GaAs Timepix detector in Time-over-threshold mode was performed with the use of Am-241 and Ba-133 radioisotopes. We were able to detect gamma-photons with the energy up to 160 keV. The X-ray imaging quality of GaAs Timepix detector was tested with X-ray source using various samples. After flat field we obtained very promising imaging performance of tested GaAs Timepix detectors.

In this work, we have focused on Timepix detectors coupled with the semi-insulating GaAs material sensor. We used undoped bulk GaAs material with the thickness of 350 mu m. We prepared and tested four pixelated detectors with 165 mu m and 220 mu m pixel size with two versions of technology preparation, without and with wet chemically etched trenches around each pixel. We have carried out adjustment of GaAs Timepix detectors to optimize their performance. The energy calibration of one GaAs Timepix detector in Time-over-threshold mode was performed with the use of Am-241 and Ba-133 radioisotopes. We were able to detect gamma-photons with the energy up to 160 keV. The X-ray imaging quality of GaAs Timepix detector was tested with X-ray source using various samples. After flat field we obtained very promising imaging performance of tested GaAs Timepix detectors.