ÚTEFČVUT Ústav technické a experimentální fyziky ČVUT v PrazeInstitute of Experimental and Applied Physics, CTU in Prague

BrachyView: Proof-of-principle of a novel in-body gamma camera for low dose-rate prostate brachytherapy

NázevTitle
BrachyView: Proof-of-principle of a novel in-body gamma camera for low dose-rate prostate brachytherapyBrachyView: Proof-of-principle of a novel in-body gamma camera for low dose-rate prostate brachytherapy
Druh výsledkuResult type
Článek v časopiseJournal article
AutořiAuthors
M. Petasecca, K. Loo, M. Safavi-Naeini, Z. Han, P. E. Metcalfe, S. Meikle, S. Pospíšil, J. Jakůbek, J. A. Bucci, M. Zaider, M. L. Lerch, Y. Qi, A. B. Rosenfeld
Klíčová slovaKeywords
Brachytherapy, prostate cancer, pixel detector, Timepix
DOIDOI
10.1118/1.4794487
Časopis / citaceJournal / citation
Medical Physics 40(4), 041709 (2013) · ISSN 0094-2405
RokYear
2013
JazykLanguage
eng
ZáznamyRecords
ProjektProject
Pokročilé techniky detekce ionizujícího zářeníAdvanced Techniques for Detection of Ionizing Radiation
CitovánoCited by
19 (OpenAlex)
2013: 22014: 22015: 22016: 22017: 32018: 22019: 32020: 12021: 02022: 02023: 02024: 2
Citace ke staženíDownload citation
TXT · BibTeX

AbstraktAbstract

BrachyView incorporates three Medipix2 detectors coupled to a multipinhole collimator. Three-dimensionally triangulated seed positions from multiple planar images are used to determine the seed placement in a PMMA prostate phantom in real time. MATLAB codes were used to test the reconstruction method and to optimize the device geometry.

BrachyView incorporates three Medipix2 detectors coupled to a multipinhole collimator. Three-dimensionally triangulated seed positions from multiple planar images are used to determine the seed placement in a PMMA prostate phantom in real time. MATLAB codes were used to test the reconstruction method and to optimize the device geometry.